14
14-10 z.one ultra and z.one ultra sp OPERATOR MANUAL
Figure 14-16. Restore Presets | Single Preset Screen
a. Select Transducer, exam type (Category), and user-created Preset, e.g.
User2 in the figure. Click Apply. The following message displays:
Figure 14-17. Overwrite Prompt
b. Select Yes . (Selecting No aborts the restore.) The system restores the
Preset. You may select another transducer/exam to restore or click Cancel
to return to imaging.
System To Back Up System
1. Power on the SmartCart/SmartCart sp or Scan Engine and allow it to fully
boot.
2. Insert USB stick for backup/restore into USB port at rear of SmartCart/
SmartCart sp
or Scan Engine.
3. A popup window will appear saying Install media detected. Do you wish to
install new software? Select No and press the Set button.
4. Press Setup button.
5. Highlight and select Preset Admin (see Figure 14-6).
6. Highlight and select Backup System (see Figure 14-7). The Backup
Configuration screen displays.