7
z.one ultra and z.one ultra sp OPERATOR MANUAL 7-69
Compounding
Scan Engine
■ Press Compounding softkey to toggle
Compounding ON/OFF.
■ Display Imaging menu and toggle
Compounding ON/OFF
Strain/Norm Strain/
Cross Corr
Scan Engine
■ Rotate Cross Corr softkey to select
desired strain acquisition format; then
press knob to activate selection.
■ Display Imaging menu and select
desired strain acquisition format.
Gain ✱
Scan Engine
■ Rotate Gain softkey to control the level of
estimated strain range displayed.
■ Rotate Gain button to adjust Gain.
VA/LIN
Scan Engine
■ Press VA/LIN softkey to switch between
LIN and VA.
■ Display Imaging menu and toggle VA/
LIN.
Scale
Scan Engine
■ Press Scale softkey to select the range
(S2 - S20) of hard/soft structures in the
image.
NOTE: For tissues with large differences in
strain, use a higher Scale. For tissues with
small differences in strain, use a lower
Scale.
■ Display Imaging menu and select range
(S2 - S20) of hard/soft structures in
image.
Table 7-7. Elastography User Controls (Continued)
Control SmartCart & SmartCart sp Scan Engine