13
13-16 z.one ultra and z.one ultra sp OPERATOR MANUAL
ICMP Echo: xx.xx.xx.xx timed out [xx.xx.xx.xx is the IP address of
the configured device].
b. Select Verify.
If a DICOM-level connection has been made, the following message
appears:
Verification of communication succeeded
If a connection has not been made, the message Connection failure
displays. If so, repeat this procedure from Step 1 or contact your
network/
PACS administrator.
7. To enable the Destination, go to the DICOM MPPS Administration screen
(
Figure 13-13), select the Destination and click Enable
8. To exit, press Enter (Enter/Select for SmartCart sp) or Freeze.
DICOM Worklist
If a DICOM Worklist application is installed on the network, a list of scheduled
exams can be delivered to the ZONARE ultrasound system on demand. One or
more worklists can be configured, but only one can be designated as the active
worklist at any given time. The worklist can be configured to specify a date range
and other criteria.
To Configure DICOM Worklist
1. Make sure the System has a live Ethernet connection to the target DICOM
devices (
Figure 13-7).
NOTE: DICOM parameters can be configured while disconnected from the
Ethernet connection, but testing will not be possible.
Table 13-5. DICOM MPPS Destination Parameters
Parameter Description
Nickname
e.g., MPPS Server
Informal name to identify the device to users in
the department. This name is typically
assigned in the department.
Application Title
e.g., MPPS_1
Name assigned to the target DICOM MPPS
Server device by the PACS Admin to identify it
on the DICOM network.
IP Address TCP/IP network IP address assigned by the IT
Admin, to the target DICOM MPPS Server
device.
Port Number TCP/IP port number assigned by the PACS
Admin, for use by the DICOM MPPS Server
device. Default port value is 104.
Max PDU Receive/Send Size Size (in bytes) of the basic data packets (PDU)
that will be transferred during receive and
transmit operations (default is 16384 = 16K).